Simulating the secondary electron avalanche of MCP by Geant4

Huaxing Peng, Baojun Yan,Han Miao, Shulin Liu,Binting Zhang

NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT(2024)

引用 0|浏览1
暂无评分
摘要
Nowadays, Microchannel Plate (MCP), as a kind of electron multipliers based on the secondary electron emission, is widely used in many high -sensitive experiments, such as neutrino detection, which require the noise to be as low as possible, while the conventional straight-channel MCP will inevitably have ion feedback, resulting in the sequential after-pulses being the major source of noise. Normally, the problem can be effectively avoided by coupling two straight-channel MCPs in cascade and combining the channels into a "V"shape known as chevron MCPs, but this method is limited by the manufacturing techniques due to the unavoidable gap between the two pieces that will worsen the resolution and peak-to-valley ratio. However, the ion feedback can be inhibited significantly for MCPs with curved channels. Based on Geant4, we investigate how the geometrical parameters of curved-channel MCP influence the gain and get the optimum pore diameter for an MCP to reach the maximum gain with fixed thickness and applied voltage. Additionally, the track-by-track simulation reveals that the average acceleration distance of a secondary electron inside the curved-channel is approximately 20 mu m when the applied voltage, length-to-diameter ratio and pore diameter are 950 V, 50:1 and 20 mu m, respectively.
更多
查看译文
关键词
MCP,Geant4,Monte Carlo,Furman model Secondary electron emission