Magnetic particle image scanner based on asymmetric core-filled electromagnetic actuator

COMPUTERS IN BIOLOGY AND MEDICINE(2024)

引用 0|浏览2
暂无评分
摘要
Monitoring the distribution of magnetic nanoparticles (MNPs) in the vascular system is an important task for the advancement of precision therapeutics and drug delivery. Despite active targeting using active motilities, it is required to visualize the position and concentration of carriers that reach the target, to promote the development of this technology. In this work, a feasibility study is presented on a tomographic scanner that allows monitoring of the injected carriers quantitatively in a relatively short interval. The device is based on a small-animal-scale asymmetric magnetic platform integrated with magnetic particle imaging technology. An optimized isotropic field-free region (FFR) generation method using a magnetic manipulation system (MMS) is derived and numerically investigated. The in-vitro and in-vivo tracking performances are demonstrated with a high position accuracy of approximately 1 mm. A newly proposed tracking method was developed, specialized in vascular system, with quick scanning time (about 1s). In this paper, the primary function of the proposed system is to track magnetic particles using a magnetic manipulation system. Through this, proposed method enables the conventional magnetic actuation systems to upgrade the functionalities of both manipulation and localization of magnetic objects.
更多
查看译文
关键词
Magnetic particle imaging,Asymmetric magnetic system,Microrobot localization and manipulation