Ex vivo biomechanical evaluation of 2.4 mm LCP plate rod constructs versus 2.7 mm LCP applied to the feline tibia

Andreas Gutbrod,Federico Longo, Roman Affentranger, Stephen J. Ferguson,Antonio Pozzi,Sebastian Knell

Veterinary Surgery(2023)

引用 0|浏览1
暂无评分
摘要
Abstract Objective To compare the stiffness and strength of three plate and rod fixation constructs applied to a feline tibial gap model. Study design Ex vivo study. Sample population Thirty‐three unpaired tibiae obtained from skeletally mature cats. Methods The tibiae were randomly divided into three groups. The following implants were then applied to the feline tibiae prior to the creation of a 10 mm diaphyseal gap. Group 1: 2.4 mm locking compression plate (LCP) and 1.0 mm intramedullary pin (IMP). Group 2: 2.4 mm LCP and 1.6 mm IMP. Group 3: 2.7 mm LCP. Subsequently, each specimen was tested for torsion, axial compression, and axial load until construct failure. Student's t ‐tests were used to compare the torsional and axial stiffness, yield load, and maximum axial force. Results Group 2 had higher axial stiffness than group 3 ( p = .013). Group 1 showed a lower maximum axial force and yield point than groups 2 and 3 ( p < .01; p < .05, respectively). There were no among‐group differences in torsional stiffness. Conclusion Constructs with a 2.4 mm LCP and 1.6 mm IMP provided the strongest and most rigid constructs in a feline tibia gap model. Clinical significance A plate‐rod construct combining a 2.4 mm LCP and a 1.6 mm IMP is appropriate for achieving high implant stiffness and resisting maximum axial force in treatment of tibial fractures in cats.
更多
查看译文
关键词
biomechanical evaluation,plate rod constructs
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要