Impact of a handpiece with a built-in fluidics pressure sensor on phacoemulsification: a multicentre prospective comparative study

Antoine P. Brezin,Dominique Monnet,Francois Lignereux, Pascal Rozot, Lea Jilet, Hendy Abdoul

BMJ Open Ophthalmology(2023)

引用 0|浏览1
暂无评分
摘要
Objective: We assessed whether the immediate pressure adjustments provided by a phacoemulsification handpiece with a built-in pressure sensor (Active Sentry, AS) could decrease the amount of energy delivered in the eye during cataract surgery.Methods and analysis: The Study of Active Sentry in Cataract Surgery was a multicentre prospective clinical trial. The Centurion system (Alcon Labs) was used and we compared phacoemulsifications performed with the AS handpiece with procedures using a sensor placed in the phacoemulsifier (non-AS). The primary outcome measure was the cumulative dissipated energy (CDE) used during the procedures. The secondary outcome measures were the total, longitudinal and torsional ultrasound (US) times, the duration of the surgery and the amount of fluid used during the surgeries.Results: The study analysed 1432 (800 AS and 632 non-AS) procedures. The mean patient age was 72.4 +/- 10.0 years. The median CDE was respectively 6.3 (IQR 3.5-9.9) and 6.7 (IQR 4.5-11.5) with and without AS (p=0.0001). The median US time was, respectively, 48.0 s (IQR 34.0-68.0) and 55.0 s (IQR 38.0-80.0) with and without AS (p<0.0001). Torsional and longitudinal US times were reduced with AS. The median duration of the procedure was, respectively, 9.0 min (IQR 7.0-12.0) and 10.0 min (IQR 7.0-13.0) (p=0.0002) with and without AS. The median balanced salt solution volume used was 52.0 cm(3) (IQR 41.0-72.0) and 57.0 cm(3) (IQR 42.0-81.0) with and without AS (p=0.0018).Conclusion: The phacoemulsifications performed with a pressure sensor built in the handpiece delivered less energy in the eye and were shorter.
更多
查看译文
关键词
Treatment Surgery, Clinical Trial, Cataract
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要