Quantification performance of silicon photomultiplier-based PET for small 18F-, 68Ga- and 124I-avid lesions in the context of radionuclide therapy planning.

Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB)(2023)

引用 0|浏览2
暂无评分
摘要
PURPOSE:The aim of this study was to investigate conditions for reliable quantification of sub-centimeter lesions with low18F,68Ga, and124I uptake using a silicon photomultiplier-based PET/CT system. METHODS:A small tumor phantom was investigated under challenging but clinically realistic conditions resembling prostate and thyroid cancer lymph node metastases (6 spheres with 3.7-9.7 mm in diameter, 9 different activity concentrations ranging from about 0.25-25 kBq/mL, and a signal-to-background ratio of 20). Radionuclides with different positron branching ratios and prompt gamma coincidence contributions were investigated. Maximum-, contour-, and oversize-based partial volume effect (PVE) correction approaches were applied. Detection and quantification performance were estimated, considering a ±30 % deviation between imaged-derived and true activity concentrations as acceptable. A standard and a prolonged acquisition time and two image reconstruction algorithms (time-of-flight with/without point spread function modelling) were analyzed. Clinical data were evaluated to assess agreement of PVE-correction approaches indicating lesion quantification validity. RESULTS:The smallest 3.7-mm sphere was not visible. If the lesions were clearly observed, quantification was, except for a few cases, acceptable using contour- or oversized-based PVE-corrections. Quantification accuracy did not substantially differ between 18F, 68Ga, and 124I. No systematic differences between the analyzed reconstruction algorithms or shorter and larger acquisition times were observed. In the clinical evaluation of 20 lesions, an excellent statistical agreement between oversize- and contour-based PVE-corrections was observed. CONCLUSIONS:At the lower end of size (<10 mm) and activity concentration ranges of lymph-node metastases, quantification with reasonable accuracy is possible for 18F, 68Ga, and 124I, possibly allowing pre-therapeutic lesion dosimetry and individualized radionuclide therapy planning.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要