Validation of the iHealth Track upper-arm blood pressure monitor KN-550BT in general population according to the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018+AMD1:2020)

Lei Shi, Xin-Yi Zhang, Jia-Yi Sun, Jin-Feng Chen,Yi-Fei Ma,Kai-Yuan Zhou

Blood pressure monitoring(2024)

引用 0|浏览3
暂无评分
摘要
Objective To validate the iHealth Track KN-550BT oscillometric upper-arm blood pressure monitor in general population according to the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018+AMD1:2020).Methods Participants were recruited and the same left-arm sequential method was used for blood pressure measurement according to the ISO 81060-2:2018+AMD1:2020. The validation results were assessed following the protocol and the Bland-Altman scatterplot was used to show the difference between the test device and reference results.Results A total of 89 qualified participants were included in the final analysis. For the validation Criterion 1, the mean +/- SD of the differences between the test device and reference readings was -1.22 +/- 5.76 mmHg and -0.08 +/- 4.40 mmHg for systolic and diastolic blood pressure, respectively. For Criterion 2, the mean +/- SD of the differences between the test device and reference readings per participant was -1.22 +/- 5.06 mmHg and -0.08 +/- 3.84 mmHg for systolic and diastolic blood pressure, respectively.Conclusion The iHealth Track KN-550BT upper-arm blood pressure monitor passed all the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018+AMD1:2020) and can be recommended for clinical use and self-measurement in general population.
更多
查看译文
关键词
AAMI/ESH/ISO Universal Standard,blood pressure monitor,ISO 81060-2:2018+AMD1:2020,validation
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要