Comparison of whole body versus thoracic bioimpedance in relation to ultrafiltration volume and systolic blood pressure during hemodialysis

JOURNAL OF APPLIED PHYSIOLOGY(2023)

引用 0|浏览6
暂无评分
摘要
In contrast to whole body bioimpedance, which estimates fluid status at a single point in time, thoracic bioimpedance applied by a wearable device could enable continuous measurements. However, clinical experience with thoracic bioimpedance in patients on dialysis is limited. To test the reproducibility of whole body and thoracic bioimpedance measurements and to compare their relationship with hemodynamic changes during hemodialysis, these parameters were measured pre- and end-dialysis in 54 patients during two sessions. The resistance from both bioimpedance techniques was moderately reproducible between two dialysis sessions (intraclass correlations of pre- to end-dialysis whole body and thoracic resistance between session 1 and 2 were 0.711 [0.58-0.8] and 0.723 [0.6-0.81], respectively). There was a very high to high correlation between changes in ultrafiltration volume and changes in whole body thoracic resistance. Changes in systolic blood pressure negatively correlated to both bioimpedance techniques. Although the relationship between changes in ultrafiltration volume and changes in resistance was stronger for whole body bioimpedance, the relationship with changes in blood pressure was at least comparable for thoracic measurements. These results suggest that thoracic bioimpedance, measured by a wearable device, may serve as an interesting alternative to whole body measurements for continuous hemodynamic monitoring during hemodialysis.
更多
查看译文
关键词
blood pressure,hemodialysis,thoracic bioimpedance,ultrafiltration volume,whole body bioimpedance
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要