Reliability and performance of the IRRAflow® system for intracranial lavage and evacuation of hematomas-A technical note.

PloS one(2024)

引用 0|浏览5
暂无评分
摘要
BACKGROUND:Intraventricular hemorrhage (IVH) is a severe condition with poor outcomes and high mortality. IRRAflow® (IRRAS AB) is a new technology introduced to accelerate IVH clearance by minimally invasive wash-out. The IRRAflow® system performs active and controlled intracranial irrigation and aspiration with physiological saline, while simultaneously monitoring and maintaining a stable intracranial pressure (ICP). We addressed important aspects of the device implementation and intracranial lavage. METHOD:To allow versatile investigation of multiple device parameters, we designed an ex vivo lab setup. We evaluated 1) compatibility between the IRRAflow® catheter and the Silverline f10 bolt (Spiegelberg), 2) the physiological and hydrodynamic effects of varying the IRRAflow® settings, 3) the accuracy of the IRRAflow® injection volumes, and 4) the reliability of the internal ICP monitor of the IRRAflow®. RESULTS:The IRRAflow® catheter was not compatible with Silverline bolt fixation, which was associated with leakage and obstruction. Design space exploration of IRRAflow® settings revealed that appropriate settings included irrigation rate 20 ml/h with a drainage bag height at 0 cm, irrigation rate 90 ml/h with a drainage bag height at 19 cm and irrigation rate 180 ml/h with a drainage bag height at 29 cm. We found the injection volume performed by the IRRAflow® to be stable and reliable, while the internal ICP monitor was compromised in several ways. We observed a significant mean drift difference of 3.16 mmHg (variance 0.4, p = 0.05) over a 24-hour test period with a mean 24-hour drift of 3.66 mmHg (variance 0.28) in the pressures measured by the IRRAflow® compared to 0.5 mmHg (variance 1.12) in the Raumedic measured pressures. CONCLUSION:Bolting of the IRRAflow® catheter using the Medtronic Silverline® bolt is not recommendable. Increased irrigation rates are recommendable followed by a decrease in drainage bag level. ICP measurement using the IRRAflow® device was unreliable and should be accompanied by a control ICP monitor device in clinical settings.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要