HPLC Method Validation for Quantification of Lisinopril

International journal of pharmaceutical quality assurance(2023)

引用 0|浏览0
暂无评分
摘要
This study aimed to develop a straightforward, sensitive, exact, quick, and accurate reverse phase high performance liquid chromatography (RP-HPLC) method for figuring out how much lisinopril is in pharmaceutical gels and other large amounts of medication. Agilent Zorbax Bonus-RP column (250 x 4.6 mm, 5μ) was used for the chromatographic separation. “A mobile phase composed of methanol and trifluoroacetic acid (50:50 v/v) was used to develop the analytical procedure. The flow was found to be occurring at a rate of 1-mL/min and with a wavelength of 215 nm. The retention time was 2.28 min. In a concentration range from 3–7 μg/mL (r2=0.998), the drug’s response was determined to be linear. The LoQ was 1.11 μg/mL, while the LoD was 0.36 μg/mL. Lisinopril’s %assay was determined to be 98.22%, while assays for the other medicines in the commercial formulation showed no interference from the excipients. This method functions well and can be applied to routine analysis.
更多
查看译文
关键词
hplc method validation,hplc method,quantification