Development of phantom for current modulation quality assurance test on computed tomography.

Gustavo E de Camargo, Glaube N Carneiro, Jessica V Real, Renato B Doro,Anna L M C Malthez

Radiation protection dosimetry(2023)

引用 0|浏览3
暂无评分
摘要
Automatic tube current modulation (ATCM) in computed tomography (CT) is a powerful tool for reducing the dose delivery to the patient during the scan. A phantom is used to perform the ATCM quality control (QC) test, assessing how the CT system adapts the tube current as a function of object size. Considering quality assurance recommendations (Brazilian and international), we developed a dedicated phantom for the ATCM test. The phantom was built-in high-density polyethylene in a cylindric shape with three different sizes. To verify the applicability of this phantom, we tested it in two different CT scanners (Toshiba and Philips). The results demonstrated concordance between a discrete change in phantom size, and the corresponding change in the tube current, indicating that the CT system could adapt current when a discrete attenuation shift occurred. The developed phantom might be used in ATCM QC tests.
更多
查看译文
关键词
computed tomography,phantom,quality assurance