Comparison of derived correction factors for effects of ion recombination and photon beam quality index following TG-51 and TRS-398 dosimetry protocols.

Seyed Mehdi Hosseini Bojdani,Hamid Reza Baghani,Mostafa Robatjazi, Stefano Andreoli,Behnam Azadegan

Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine(2023)

引用 0|浏览0
暂无评分
摘要
In this study, ion recombination correction factor (k) and beam quality conversion factor ( [Formula: see text] ) values were extracted following the recommendations of the TRS-398 and TG-51 dosimetry protocols for widely used cylindrical ionization chambers for high energy photon beam dosimetry to quantify the agreement between the instructions for these two protocols for absolute dosimetry inside water. Four different types of cylindrical ionization chambers comprising Farmer (TM30013), Semiflex 0.125 cm (TM31010), Semiflex 0.3 cm (TM31013), and PinPoint (TM31016) were considered, and k and [Formula: see text] values were determined at photon energies of 6 MV and 15 MV. The maximum difference between the measured k values according to the instructions in the TRS-398 and TG-51 protocols was 0.03%. The k data measured with both protocols agreed well with those measured by using the Jaffe-plot approach, where the maximum difference was about 0.33%. The observed differences between the [Formula: see text] factors measured by using the TRS-398 and TG-51 dosimetry protocols at photon energies of 6 MV and 15 MV were 0.37% and 0.55%, respectively. The [Formula: see text] values measured using the TG-51 dosimetry protocols were slightly closer to those measured by a reference ionization chamber dosimeter. We conclude that the maximum differences were about 0.4% and 0.6% in the absorbed dose measurements according to the TRS-398 and TG-51 instructions at photon energies of 6 MV and 15 MV, respectively. The type of ionization chamber employed also affected the differences, where the maximum and minimum dose differences were found using the Farmer and PinPoint chambers, respectively.
更多
查看译文
关键词
Beam quality conversion Factor,Dosimetry,High energy photon beam,Influence quantity,Ion recombination correction factor
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要