Total scattering measurements at the Australian Synchrotron Powder Diffraction beamline: capabilities and limitations

Journal of synchrotron radiation(2023)

引用 0|浏览26
暂无评分
摘要
This study describes the capabilities and limitations of carrying out total scattering experiments on the Powder Diffraction (PD) beamline at the Australian Synchrotron, ANSTO. A maximum instrument momentum transfer of 19 A degrees similar to 1 can be achieved if the data are collected at 21 keV. The results detail how the pair distribution function (PDF) is affected by Qmax, absorption and counting time duration at the PD beamline, and refined structural parameters exemplify how the PDF is affected by these parameters. There are considerations when performing total scattering experiments at the PD beamline, including (1) samples need to be stable during data collection, (2) highly absorbing samples with a similar to R > 1 always require dilution and (3) only correlation length differences > 0.35 A degrees may be resolved. A case study comparing the PDFatom-atom correlation lengths with EXAFS-derived radial distances of Ni and Pt nanocrystals is also presented, which shows good agreement between the two techniques. The results here can be used as a guide for researchers considering total scattering experiments at the PD beamline or similarly setup beamlines.
更多
查看译文
关键词
pair distribution function,powder X-ray diffraction,Mythen-II detector,total scattering
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper