Development of 3 He-filled linear position-sensitive detector for neutron scattering instruments at CSNS

Radiation Detection Technology and Methods(2022)

引用 0|浏览11
暂无评分
摘要
Background Helium-3-filled linear position-sensitive detectors (LPSDs) have been widely applied to neutron scattering instruments in recent decades owing to the high detection efficiency, the excellent neutron/gamma discrimination, and the ability to construct the detector with large area coverage. More than 65% of neutron instruments at the China Spallation Neutron Source (CSNS) require 3 He LPSDs for the building of the detector system. The detector of a neutron scattering instrument is normally an array detector composed of a large number of 3 He LPSDs. However, no appropriate substitute detector is available for the 3 He LPSD for operation over a short period, and its specifications need to be customized to satisfy the requirements of different instruments. This necessitates the development of 3 He LPSDs with different specifications for building and upgrading the detectors of neutron scattering instruments at CSNS. Purpose To meet the general requirements specified by the neutron instruments at CSNS, a 3 He LPSD was developed and its performances were investigated by using the neutron beam. Methods The neutron beam experiments for this 3 He LPSD, including measurements of counting rate plateau curve, position calibration, and position resolution along the tube path, were carried out at the beamline-20 of CSNS. To evaluate its performance in the instrument operation, a detector module made of this 3 He LPSD has been installed in the multi-physics instrument (MPI), and the comparison test with the commercial 3 He LPSD has been conducted at neutron scattering experiments. Results and conclusion The experimental measurements for this 3 He LPSD based on the neutron beam showed that it satisfies the requirements for use in neutron scattering instruments. It reaches a plateau slope of 2.6%/100 V with a plateau range from 1750 to 1950 V and attained a minimal position resolution of 5.3 mm (FWHM), superior to the one achieved by the commercial 3 He LPSD. In addition, this 3 He LPSD detector module applied to the MPI has been working steadily for half of year and is available for use in experiments. These results provide a sound basis for the subsequent construction and upgradation of the detector of neutron scattering instruments at CSNS.
更多
查看译文
关键词
Counting rate, CSNS, LPSD, Plateau curve, Position resolution
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要