Influence of different air CT numbers for IVDT on the dose distribution in TomoTherapy MVCT.

Journal of applied clinical medical physics(2023)

引用 0|浏览1
暂无评分
摘要
This study aims to evaluate the effect of different air computed tomography (CT) numbers of the image value density table (IVDT) on the retrospective dose calculation of head-and-neck (HN) radiotherapy using TomoTherapy megavoltage CT (MVCT) images. The CT numbers of the inside and outside air and each tissue-equivalent plug of the "Cheese" phantom were obtained from TomoTherapy MVCT. Two IVDTs with different air CT numbers were created and applied to MVCT images of the HN anthropomorphic phantom and recalculated by Planned Adaptive to verify dose distribution. We defined the recalculation dose with MVCT images using both inside and outside air of the IVDT as and , respectively. Treatment planning doses calculated on kVCT images were compared with those calculated on MVCT images using two different IVDT tables, namely, and . The difference between average MVCT numbers ±1 standard deviation on inside and outside air of the calibration phantom was 65 ± 36 HU. This difference in MVCT number of air exceeded the recommendation lung tolerance for dose calculation error of 2%. The dose differences between the planning target volume (PTV): D , D , D and the organ at risk (OAR): D , D recalculated by and using MVCT images were a maximum of 0.7% and 1.2%, respectively. Recalculated doses to the PTV and OAR with MVCT showed that was 0.5%-0.7% closer to the kVCT treatment planning dose than . This study showed that was more accurate than in recalculating the dose HN cases of MVCT using TomoTherapy.
更多
查看译文
关键词
ART,IVDT,MVCT,air CT number,head-and-neck radiotherapy,helical TomoTherapy
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要