Plastic Energy Dissipation in Lumbar Spine Implants: A Contact Mechanics Point of View

Journal of Engineering and Science in Medical Diagnostics and Therapy(2019)

引用 0|浏览0
暂无评分
摘要
In this study, an elastoplastic contact model is developed for L1–L5 lumbar spine implants. Roughness effect is included to estimate energy loss which is an indication of wear and subsequently the issue of metal debris in body. A Gaussian function is assumed for the distribution of asperities. The contact surfaces of the implants are assumed to be spherical caps. Subsequently, a least-square approach is applied to obtain an approximate expression for the contact force using the data from integration over contact zone. The energy loss is calculated, next, which is due to plastic deformations of asperities. The numerical results indicate that for a given loading–unloading condition, the amount of energy dissipation increases in L1–L4 lumbar spine implants, while it decreases from L4 to L5 implants. The implants geometrical specifications are chosen to cover a wide range of patients' age. Finally, a closed-form expression is obtained for the plastic energy dissipation per cycle in terms of plasticity index for the lumbar spine L4, as the worst-case scenario. Such a function can serve as a very useful tool for implant designers and manufacturers.