Use of a novel pedal-operated compressor is non-inferior to the use of a standard hand-compressed bag-valve mask

JOURNAL OF THE AMERICAN COLLEGE OF EMERGENCY PHYSICIANS OPEN(2022)

引用 1|浏览0
暂无评分
摘要
Background The standard bag-valve mask (BVM) used universally requires that a single healthcare practitioner affix the mask to the face with 1 hand while compressing a self-inflating bag with the second hand. Studies have demonstrated that creating a 2-handed seal (with 2 healthcare practitioners) is superior. Our study aims to assess the efficacy of a novel single-practitioner BVM device that uses a foot pedal as the bag compressor, allowing both hands to be available for the seal to facilitate delivery of appropriate tidal volumes during single-practitioner resuscitation. Methods This was a prospective, randomized, cross-over study. Participants with various BVM ventilation experience performed 2 minutes of metronome-guided BVM ventilation using a standard BVM and the pedal-operated compressor on a high-fidelity simulation mannequin. Analysis examining differences in mean tidal volume delivered was conducted using a regression model that adjusted for covariates. A secondary analysis using a series of Wilcoxon tests was conducted to compare differences in the additional out-of-range sensed breaths metrics to compare differences by prior BVM ventilation experience. Results A total of 58 subjects participated. The pedal-operated compressor unadjusted mean tidal volume delivered was 446.5 mL (95% confidence interval [CI], 425.9-467.1) compared with 340.6 mL (95% CI, 312.2-369.0) by standard BVM (mean change, 105.9 mL [95% CI, 71.2-140.6]; P < .001). When modeling a generalized estimation equation regression model, standard BVM ventilation provided a mean difference of 105.9 mL less than pedal-operated compressor ventilation after adjusting for covariates (P = 0.01). For the secondary outcome, the pedal-operated compressor did have a significantly lower median number of out-of-range breaths (median, 3; interquartile range [IQR], 1-11.5) compared with the standard device (median, 13.5; IQR, 6-19; P < 0.001). Conclusions Use of a novel pedal-operated compressor may allow a single healthcare practitioner, regardless of prior experience, to deliver consistent, appropriate tidal volumes with more ease compared with the standard BVM during manual respiratory resuscitation.
更多
查看译文
关键词
critical care, emergency medicine, inventions, resuscitation, tidal volume, ventilation
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要