The PRONTO study: Clinical performance of ID NOW in individuals with compatible SARS-CoV-2 symptoms in walk-in centres-accelerated turnaround time for contact tracing.

Canada communicable disease report = Releve des maladies transmissibles au Canada(2021)

引用 4|浏览2
暂无评分
摘要
Background:This PRONTO study investigated the clinical performance of the Abbott ID NOWTM (IDN) COVID-19 diagnostic assay used at point of care and its impact on turnaround time for divulgation of test results. Methods:Prospective study conducted from December 2020 to February 2021 in acute symptomatic participants presenting in three walk-in centres in the province of Québec. Results:Valid paired samples were obtained from 2,372 participants. A positive result on either the IDN or the standard-of-care nucleic acid amplification test (SOC-NAAT) was obtained in 423 participants (prevalence of 17.8%). Overall sensitivity of IDN and SOC-NAAT were 96.4% (95% CI: 94.2-98.0%) and 99.1% (95% CI: 97.6-99.8), respectively; negative predictive values were 99.2% (95% CI: 98.7-99.6%) and 99.8% (95% CI: 99.5-100%), respectively. Turnaround time for positive results was significantly faster on IDN. Conclusion:In our experience, IDN use in symptomatic individuals in walk-in centres is a reliable sensitive alternative to SOC-NAAT without the need for subsequent confirmation of negative results. Such deployment can accelerate contact tracing, reduce the burden on laboratories and increase access to testing.
更多
查看译文
关键词
Abbott ID NOW,COVID-19,Canada,SARS-CoV-2,diagnostic performance,nucleic acid amplification tests,point-of-care testing,predictive value,rapid tests,sensitivity and specificity