Design of SPECT for BNCT to measure local boron dose with GAGG scintillator

Applied Radiation and Isotopes(2022)

引用 8|浏览11
暂无评分
摘要
Boron Neutron Capture Therapy (BNCT) is a promising cancer therapy, which has recently been in practical use in Japan especially using an accelerator. In BNCT real-time measurement of local boron dose is required. In the present study, a novel design of a SPECT system for BNCT (BNCT-SPECT) has been carried out to realize estimation of the local boron dose, i.e., treatment effect of BNCT. Necessary performance which BNCT community requires includes accuracy of 5% and spatial resolution of 5 mm, which are regarded to be difficult to realize. A possible design was investigated to meet the requirements. The design results we achieved are as follows: As for the elemental detection device, GAGG (3.5 × 3.5 × 30 mm3) was selected, and for the collimator, the collimator hole diameter was 3.5 mm, the collimator hole pitch was 4 mm and the collimator length was 26 cm. For the obtained performance with the design, the accuracy was 4.4% and the spatial resolution was 5.1 mm. Currently prior to production of the real system, a prototype of BNCT-SPECT is being developed to acquire real projection data to confirm the performance and examine our own image reconstruction method with the obtained projection data.
更多
查看译文
关键词
local boron dose,bnct,gagg scintillator,spect