Contrast-enhanced Ultrasound Imaging Using Capacitive Micromachined Ultrasonic Transducers

JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA(2021)

引用 1|浏览0
暂无评分
摘要
Capacitive micromachined ultrasonic transducers (CMUTs) have a nonlinear relationship between the applied voltage and the emitted signal, which is detrimental to conventional contrast-enhanced ultrasound (CEUS) techniques. Instead, a three-pulse amplitude modulation (AM) sequence has been proposed, which is not adversely affectedby the emitted harmonics. In this paper, this is shown theoretically, and the performance of the sequence is verified using a 4.8 MHz linear CMUT array, and a comparable lead zirconate titanate (PZT) array, across 6 V to 60 V applied AC voltage. CEUS images of the contrast agent SonoVue flowing through a 3D printed hydrogel phantom showed an average enhancement in contrast-to-tissue ratio (CTR) between B-mode and CEUS images of 49.9 dB and 37.4 dB for the PZT array and CMUT, respectively. Furthermore, hydrophone recordings of the emitted signals showed that the nonlinear emissions from the CMUT did not significantly degrade the cancellation in the compounded AM signal, leaving an average of 2% of the emitted power between 26 V and 60 V AC. Thus, it is demonstrated that CMUTs are capable of CEUS imaging independent of the applied excitation voltage when using a three-pulse AM sequence.
更多
查看译文
关键词
capacitive micromachined ultrasonic transducers,ultrasound imaging,contrast-enhanced