Optimizing Tip Diameter in Phacoemulsification of Varying Lens Sizes: An in vitro Study

CLINICAL OPHTHALMOLOGY(2021)

引用 3|浏览3
暂无评分
摘要
Purpose: We assessed the effect of two lens cube sizes, three tip sizes, and two ultrasound (US) approaches on phacoemulsification efficiency and chatter. Methods: After porcine lens nuclei were soaked in formalin, we divided them into cubes measuring 2.0 mm or 3.0 mm. We collected efficiency and chatter data for 30-degree bent 19 G, 20 G, and 21 G tips with a continuous torsional US system; and for straight 19 G, 20 G, and 21 G tips with a micropulse longitudinal US system. Results: The average time needed for removal was always higher for the 3.0 mm lens cube than for the 2.0 mm lens cube. Statistically significant differences were observed between the 19 G and 21 G tips with micropulse longitudinal US using a 2.0 mm cube and a 3.0 mm cube, and with continuous transversal US using a 2.0 mm lens cube and a 3.0 mm cube. We did not observe significant differences between 19 G and 20 G tips with either cube size in either US system. However, we noted identical trends for both cube sizes with both US approaches; 19 G tips performed better than 20 G and 21 G tips. Conclusion: Regardless of the lens size, 19 G needles were the most efficient, and had both the fewest outliers and the smallest standard deviations.
更多
查看译文
关键词
cataract, chatter, efficiency, tip diameter, vacuum