Analytical and Clinical Considerations in Implementing the Roche Elecsys Troponin T Gen 5 STAT Assay

AMERICAN JOURNAL OF CLINICAL PATHOLOGY(2021)

引用 4|浏览0
暂无评分
摘要
Objectives: To evaluate the analytical and clinical performance characteristics of the fifth-generation troponin T reagent. Methods: Troponin T was measured in 2,332 paired serum and plasma samples from emergency department and hospital patients using the fourth- and fifth-generation reagents. Testing was repeated after recentrifugation to determine the frequency of analytical outliers and percentage of patients with elevated values for each assay. We conducted separate experiments to determine the effects of biotin and hemolysis interference, as well as measure interinstrument variability, for fifth-generation troponin T. Results: Analytic outliers occurred more frequently using the fifth-generation reagent (3.4%) compared with the fourth-generation reagent (1.0%). The frequency of elevated troponin T above the 99th percentile upper reference limit was 26% for the fourth-generation reagent and 52% for the fifth-generation reagent. Clinically significant assay interference by biotin was observed at 20 ng/mL, but hemolysis interference was not observed until an H index of 150. Instrument-to-instrument variability between e411 and e601/602 instrument platforms is predicted to confound clinical interpretation of troponin changes. Conclusions: Analytical outliers and instrument-toinstrument variability are the two analytical variables most likely to confound interpretation of changes in fifthgeneration troponin T results over time.
更多
查看译文
关键词
Cardiac troponin, Analytical outliers, Analytical validation, Clinical validation
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要