Diagnosis And Monitoring Of Hepatitis B Virus Infection Using The Cobas(R) Hbv Test For Use On The Cobas(R) 4800 System

MICROORGANISMS(2021)

引用 2|浏览1
暂无评分
摘要
(1) Background: Sensitive and accurate nucleic acid amplification technologies are now recommended for hepatitis B virus (HBV) DNA detection and quantification in clinical practice to diagnose and monitor hepatitis B infection. The aim of this study was to assess the analytical and clinical performance of the cobas(R) HBV Test on the cobas(R) 4800 System. (2) Methods: Standard panel and clinical specimens were tested in parallel with three different real-time commercial PCR assays including the cobas (R) HBV Test, the Cobas(R) AmpliPrep/Cobas(R) TaqMan HBV Test v2.0 and Alinity (TM) m HBV assay. (3) Results: The specificity of the cobas(R) HBV Test was 97.9%. The limit of detection was estimated to be 2.1 IU/mL. Intra-assay and interassay coefficients of variation varied from 0.14% to 1.92% and 2.16% to 12.02%, respectively. HBV DNA levels in patients infected with different HBV genotypes strongly correlated with those measured by the two other commercial comparators assays. (4) Conclusions: The cobas(R) HBV Test can be confidently used to detect and accurately quantify HBV DNA in clinical practice as well as in clinical trials with the new anti-HBV drugs currently in development.
更多
查看译文
关键词
hepatitis B virus, HBV DNA, HBV monitoring, real-time PCR, hepatitis B infection diagnosis
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要