Diagnostic and analytical performance evaluation of ten commercial assays for detecting SARS-CoV-2 humoral immune response.

Journal of immunological methods(2021)

引用 8|浏览1
暂无评分
摘要
OBJECTIVE:Analytical validation of newly released SARS-CoV-2 antibody assays in the clinical laboratory is crucial to ensure sufficient performance in respect to its intended use. We aimed to assess analytical and diagnostic performance of 8 (semi-)quantitative assays detecting anti-nucleocapsid IgG (Euroimmun, Id-Vet) or total Ig (Roche), anti-spike protein IgG (Euroimmun, Theradiag, DiaSorin, Thermo Fisher) or both (Theradiag) and 2 rapid lateral flow assays (LFA) (AAZ-LMB and Theradiag). METHODS:Specificity was evaluated using a cross-reactivity panel of 85 pre-pandemic serum samples. Sensitivity was determined at both the manufacturer's and a 95% specificity cut-off level, using 81 serum samples of patients with a positive rRT-PCR. Sensitivity was determined in function of time post symptoms onset. RESULTS:Specificity for all assays ranged from 92.9% to 100% (Roche and Thermo Fisher) with the exception of the Theradiag IgM LFA (82.4%). Sensitivity in asymptomatic patients ranged between 41.7% and 58.3%. Sensitivity on samples taken <10 days since symptom onset was low (23.3%-66.7%) and increased on samples taken between 10 and 20 days and > 20 days since symptom onset (80%-96% and 92.9%-100%, respectively). From 20 days after symptom onset, the Roche, Id-vet and Thermo Fisher assays all met the sensitivity (>95%) and specificity (>97%) targets determined by the WHO. Antibody signal response was significantly higher in the critically ill patient group. CONCLUSION:Antibody detection can complement rRT-PCR for the diagnosis of COVID-19, especially in the later stage, or in asymptomatic patients for epidemiological purposes. Addition of IgM in LFAs did not improve sensitivity.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要