Diagnostic Accuracy Of Nasopharyngeal Swab, Nasal Swab And Saliva Swab Samples For The Detection Of Sars-Cov-2 Using Rt-Pcr

INFECTIOUS DISEASES(2021)

引用 24|浏览9
暂无评分
摘要
BackgroundThe current gold standard in coronavirus disease (COVID-19) diagnostics is the real-time reverse transcription-polymerase chain reaction (RT-PCR) assay for detecting severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) RNA in nasopharyngeal swab (NPS) samples. Alternatively, nasal swab (NS) or saliva swab (SS) specimens are used, although available data on test accuracy are limited. We examined the diagnostic accuracy of NPS/NS/SS samples for this purpose.MethodsTen patients were included after being tested positive for SARS-CoV-2 RT-PCR in NPS samples according to the National Institute of Infectious Disease guidelines. In comparison with this conventional diagnostic method, NPS/NS/SS samples were tested using the cobas 6800 systems RT-PCR device. To investigate the usefulness of the cobas method and the difference among sample types, the agreement and sensitivity were calculated. Five to six samples were collected over a total period of 5-6 d from each patient.ResultsFifty-seven sets of NPS/NS/SS samples were collected, of which 40 tested positive for COVID-19 by the conventional method. Overall, the concordance rates using the conventional method were 86.0%/70.2%/54.4% for NPS/NS/SS samples (cobas); however, for samples collected up to and including on Day 9 after disease onset (22 negative and one positive specimens), the corresponding rates were 95.7%/87.0%/65.2%. The overall sensitivity estimates were 100.0%/67.5%/37.5% for NPS/NS/SS samples (cobas). For samples up to 9 d after onset, the corresponding values were 100.0%/86.4%/63.6%.ConclusionsNS samples are more reliable than SS samples and can be an alternative to NPS samples. They can be a useful diagnostic method in the future.
更多
查看译文
关键词
COVID-19 diagnostic test, nasopharyngeal swab, nasal swab, saliva
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要