Decreased stress shielding with a PEEK femoral total knee prosthesis measured in validated computational models

Journal of Biomechanics(2021)

引用 20|浏览1
暂无评分
摘要
Due to their high stiffness, metal femoral implants in total knee arthroplasty may cause stress shielding of the peri-prosthetic bone, which can lead to loss of bone stock. Using a polymer (PEEK) femoral implant reduces the stiffness mismatch between implant and bone, and therefore has the potential to decrease strain shielding. The goal of the current study was to evaluate this potential benefit of PEEK femoral components in cadaveric experiments. Cadaveric femurs were loaded in a materials testing device, while a 3-D digital image correlation set-up captured strains on the surface of the intact femurs and femurs implanted with PEEK and CoCr components. These experimental results were used to validate specimen-specific finite element models, which subsequently were used to assess the effect of metal and PEEK femoral components on the bone strain energy density. The finite element models showed strain maps that were highly comparable to the experimental measurements. The PEEK implant increased strain energy density, relative to the preoperative bone and compared to CoCr. This was most pronounced in the regions directly under the implant and near load contact sites. These data confirm the hypothesis that a PEEK femoral implant can reduce peri-prosthetic stress shielding.
更多
查看译文
关键词
Total knee arthroplasty,Polyetheretherketone,Stress shielding,Finite element simulation,Digital image correlation