Evaluation Of A New Spinal Surgical Robotic System Of Kirschner Wire Placement For Lumbar Fusion: A Multi-Centre, Randomised Controlled Clinical Study

INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY(2021)

引用 7|浏览21
暂无评分
摘要
Background To introduce a novel robotic system 'Orthbot' that has been developed and tested as a surgical assistant for auto-placement of the K-wire in lumbar fusion.Methods This is a multi-centre, randomized controlled clinical study that includes 56 patients (robot group, RG: 27, free-hand group, FG: 29). Following the pre-operative planning and intra-operative fluoroscopic images, the 'Orthbot' automatically completed registration and K-wire placement under the supervision of the surgeon. Deviation distance (DD) and deviation angle (DA) were used as the primary parameters to evaluate the accuracy of the robotic system.Results The average DD was 0.95 +/- 0.377 mm and 4.35 +/- 2.01 mm, respectively in the RG and FG (p < 0.001). The average DA of the K-wire in the coronal plane and the sagittal plane in X-Ray was respectively 6.80 +/- 7.79 degrees and 1.27 +/- 2.32 degrees in the RG (p < 0.001), and 22.22 +/- 16.85 degrees and 4.57 +/- 3.86 degrees in the FG (p < 0.001), which showed a higher accuracy rate in the robotic-assisted cases compared to the free-hand cases.Conclusions The novel robotic system could achieve accurate K-wire insertions as indicated by the radiological results.
更多
查看译文
关键词
free&#8208, hand technique, posterior lumbar inter&#8208, body fusion, proximal facet joint violation, randomized controlled trial, robot&#8208, assisted K&#8208, wire insertion