Vuv Transmission Of Ptfe For Xenon-Based Particle Detectors

JOURNAL OF INSTRUMENTATION(2020)

引用 1|浏览5
暂无评分
摘要
Liquid xenon (LXe) based detectors for rare event searches in particle and astroparticle physics are optimized for high xenon scintillation light collection and low background rate from detector materials. Polytetrafluoroethylene (PTFE, Teflon (TM)) is commonly used to encapsulate the active LXe volume due to its high reflectance for VUV LXe scintillation light with peak emission at 178 nm Reflectance, transmission and number of background signals arising from PT FE depend on the thickness of the PTFE detector walls. In this work, we present VUV transmission measurements for PTFE of various thicknesses often considered in the design phase of LXe detectors. PTFE samples are measured in an apparatus previously used for reflectance measurements in LXe using collimated light at a wavelength of 178 nm Measurements in vacuum as well as gaseous xenon are described by the Kubelka and Munk model, as well as by Beer-Lambert's law for samples of >= 0.7 mm thickness, yielding a transmission coefficient of lambda(BL) = ( 0.89 +/- 0.05) mm The PTFE wall thickness of the XENONnT dark matter experiment was optimized by these measurements and selected as >= 3 mm.
更多
查看译文
关键词
Noble liquid detectors (scintillation, ionization, double-phase), Detector design and construction technologies and materials, Time projection Chambers (TPC), Dark Matter detectors (WIMPs, axions, etc.)