Evaluation Of N8-Gp Activity Using A One-Stage Clotting Assay: A Single-Center Experience

ACTA HAEMATOLOGICA(2020)

引用 2|浏览12
暂无评分
摘要
N8-GP (ESPEROCT (R); turoctocog alfa pegol; Novo Nordisk A/S, Bagsvaerd, Denmark) is an extended half-life recombinant factor VIII (FVIII) molecule. FVIII-deficient plasma spiked with N8-GP can be accurately measured using many activated partial thromboplastin time (aPTT)-based one-stage clotting assay reagents with normal human plasma calibrators. To date, there are few data on the measurement accuracy of samples from patients treated with N8-GP. Here, we measure patient samples during routine treatment monitoring. Three previously treated patients with severe hemophilia A (HA) without inhibitors (baseline FVIII activity <0.01 IU/mL) received 50 IU/kg N8-GP every fourth day or twice weekly over 5 years as part of the pathfinder2 trial. Patient samples were monitored using the Pathromtin (R) SL aPTT reagent (Siemens Healthcare GmbH, Erlangen, Germany), a BCS (R) XP System analyzer (Siemens), and Standard Human Plasma (Siemens) or product-specific calibrators. Patient age ranged from 36 to 62 years. Overall, measurements performed using product-specific or Standard Human Plasma calibrators were in good agreement, with ratios randomly distributed around 1.0. Peak ratios tended to be closer to 1.0 than trough samples. Pathromtin (R) SL with Standard Human Plasma calibrator consistently and accurately measured FVIII activity in samples from severe HA patients receiving N8-GP prophylaxis.
更多
查看译文
关键词
Clinical laboratory techniques, Hemophilia A, Factor VIII, Analysis, Partial thromboplastin time