A Generic, Scalable, And Cost-Effective Detector Front-End Block For Pet

2017 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC)(2017)

引用 1|浏览3
暂无评分
摘要
In this work we present a Positron Emission Tomography (PET) detector block, which can be used in different scanner geometries. The block uses 12 x 12 Silicon Photomultiplier (SiPM) pixels with the overall dimension of 50x50 mm(2). The SiPM technology has widely used in PET detectors, particularly in the presence of magnetic field. The large number of SiPM pixels per unit area, requires channel reduction technique to avoid readout complexity and cost. The presented detector block consists of the detector head and the digital front-end (DFE) board. The read-out dimension is so that the blocks can be arranged in the arbitrary numbers in a ring, or in the axial extension. In the DFE board the basic information required for a PET scanner (i.e. position, energy, and time of photon incidence), are measured.We used Scrambled Crosswire Readout (SCR) method by exploiting the third output of SiPM, which has faster rise time instead of cathode. This method reduces 144 fast outputs to 9 tile signals and 144 standard outputs to 16 energy channels. The multiplexed fast signals are also used to generate time pickoff information for timing measurement. To calculate the position of the photon, 16 energy signals locate the position inside the tile, and the tile signals locate the tile in the array. The position calculation uses threshold and centroid method which is described in detail in the tile edges, when two or four tiles are fired by a single event.The basic design uses a 24x24 LYSO:Ce array with 2x2x10 mm(3) pixel size to be used in the small animal PET project. In the 2D flood image, all the pixels are clearly resolved in the room temperature. The energy resolution of the detector block is measured similar to 12.5% by calibrating energy values for all crystal pixels.
更多
查看译文
关键词
faster rise time,multiplexed fast signals,time pickoff information,timing measurement,position calculation,centroid method,tile edges,animal PET project,crystal pixels,generic cost-effective detector front-end,Positron Emission Tomography detector block,different scanner geometries,SiPM technology,PET detectors,magnetic field,SiPM pixels,unit area,channel reduction technique,readout complexity,presented detector block,detector head,digital front-end board,arbitrary numbers,DFE board,basic information,PET scanner,Crosswire Readout method,fast outputs,tile signals,standard outputs,energy channels,energy signals,pixel size,silicon photomultiplier pixels,scalable cost-effective detector front-end,2D flood image
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要