Investigation into the limitations of straightness interferometers using a multisensor-based error separation method

MEASUREMENT SCIENCE AND TECHNOLOGY(2018)

引用 4|浏览4
暂无评分
摘要
The uncertainty of a straightness interferometer is independent of the component used to introduce the divergence angle between the two probing beams, and is limited by three main error sources, which are linked to each other: their resolution, the influence of refractive index gradients and the topography of the straightness reflector. To identify the configuration with minimal uncertainties under laboratory conditions, a fully fibre-coupled heterodyne interferometer was successively equipped with three different wedge prisms, resulting in three different divergence angles (4 degrees, 8 degrees and 20 degrees). To separate the error sources an independent reference with a smaller reproducibility is needed. Therefore, the straightness measurement capability of the Nanometer Comparator, based on a multisensor error separation method, was improved to provide measurements with a reproducibility of 0.2 nm. The comparison results revealed that the influence of the refractive index gradients of air did not increase with interspaces between the probing beams of more than 11.3 mm. Therefore, over a movement range of 220 mm, the lowest uncertainty was achieved with the largest divergence angle. The dominant uncertainty contribution arose from the mirror topography, which was additionally determined with a Fizeau interferometer. The measured topography agreed within +/- 1.3 nm with the systematic deviations revealed in the straightness comparison, resulting in an uncertainty contribution of 2.6 nm for the straightness interferometer.
更多
查看译文
关键词
heterodyne interferometer,straightness measurement,error separation
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要