Ultra-trace analysis of silver and platinum in seawater by ICP-SFMS after off-line matrix separation and pre-concentration

Marine Chemistry(2018)

引用 27|浏览5
暂无评分
摘要
Precise and accurate quantification of Ag and platinum group elements in marine, estuarine and riverine waters is challenging due to the complex matrices and ultra-low concentrations in the pmolkg−1 and fmolkg−1 range. A methodological approach has been developed for the analysis of Ag and Pt in seawater based on the extraction of their anionic chloro-complexes utilizing the strong anion exchange resin Dowex™ 1-×8 in an off-line matrix separation and pre-concentration approach prior to determination by ICP-SFMS. The method allows simultaneous processing of four 100mL samples at a flow rate of up to 2mL min−1 in an ultra-clean manifold. Enrichment factors of 40 could be achieved with elution into 2.5g of a mixture of HNO3/HCl, yielding adequately low detection limits to obtain open ocean profiles. Quantification was performed by matrix-matched calibration and internal standardization. To validate the accuracy of this approach, isotope dilution analysis was also used. The applicability of the method was demonstrated on a suite of reference samples and depth profiles collected during the 2009 2nd U.S. GEOTRACES Intercalibration cruise in the eastern North Pacific.
更多
查看译文
关键词
Trace metals,Seawater,Ultraviolet radiation,ICP-SFMS,Strong anion exchange resin