Design considerations and experimental results for MRI systems using HTS magnets

SUPERCONDUCTOR SCIENCE & TECHNOLOGY(2017)

引用 39|浏览10
暂无评分
摘要
An increasing number of magnetic resonance imaging (MRI) systems using high temperature superconductors (HTS) magnets have been designed and constructed, with detailed results of their performance now available. Features of REBCO and BSCCO conductors are described as they pertain to use in high homogeneity magnets, with emphasis placed on the practical use of these conductors in magnets. Methods of coil winding are discussed, in particular the differences between pancake and layer winding techniques. Design considerations for HTS magnets are presented in light of the difficulties presented by quench in these magnets, but also in terms of the features of HTS magnets afforded by their high operating temperatures, namely robust cryogen free operation and the potential to use unshielded gradient coils. Drawing on two example MRI systems, namely a 3 T BSCCO brain imaging magnet developed in Japan and a 1.5 T REBCO orthopaedic imaging system developed in New Zealand, the report details real-world stability and homogeneity of HTS-MRI systems, in particular with regards to the screening current effects observed in these systems. It is concluded that, apart from conductor cost, there are currently no technical obstacles to use of HTS-MRI systems.