Analytical performances of Hemoclot Protein C Reagent on ACL TOP analyzer.
ANNALES DE BIOLOGIE CLINIQUE(2016)
摘要
Our study aimed to evaluate and validate according to standard NF EN ISO 15189 the original protocol ajustement of Hemoclot Protein C (PC) (Hyphen BioMed), clotting-based assay of PC on ACL TOP analyzer (Werfen/Instrumentation Laboratory). We evaluated the performance in terms of imprecision and we validate additional parameters in range B required by the SH GTA 04 (COFRAC): repeatability, reproducibility, detection and quantification limits, limits of linearity, stability, inter-samples and inter-reagents contamination, inaccuracy, evaluation of interferences (hemolysis, bilirubinemia and chyles). A comparison with Hemoclot PC on STA Compact analyzer (Stago) was performed. Coefficients of variation were lower than 5 %. Detection and quantification limits were respectively 8.3 % and 9.3 %. Superior limit of linearity was 140 %. The test didn't diplay any inter-samples and inter-reagents contamination. Reagent after reconstitution was stable 6 hours on ACL TOP. No interferences were observed for hemoglobin lower than 500 mg/dL, for bilirubin lower than and for chyles lower than 300 mg/dL. Comparison with Hemoclot PC on STA analyzer (Stago) was satisfactory. Hemoclot PC adjusted on ACL TOP analyzer showed satisfactory analytical perfoiuiances with criteria chosen in our study. These data allow a better knowledge of the performances of this test and were useful to make a validation file in range B as recommended by SH GTA 04.