An Efficient Hardware Architecture Design of EEMD Processor for Electrocardiography Signal

2018 IEEE Biomedical Circuits and Systems Conference (BioCAS)(2018)

引用 1|浏览1
暂无评分
摘要
This study proposed an efficient hardware architecture design of Ensemble Empirical Mode Decomposition (EEMD) processor for the signal analysis of Electrocardiography (ECG). The proposed processor is implemented in an on-board Xilinx FPGA for on-line signal processing of the non-linear and non-stationary signal. The EEMD method is appropriate to analyze the non-linear ECG signal with assisting white noise and decompose the signal into 8 sets of Intrinsic Mode Functions (IMFs). The experimental result shows that the mode mixing problem, which exists in the Empirical Mode Decomposition (EMD) method, solved by the proposed EEMD processor. The study solves the obstacle of mode mixing and achieves high accuracy with data error <; 4.7×10-5. This approach can effectively analyze the non-linear and non-stationary biomedical signal and facilitate cardiovascular diseases diagnosis and long-term monitoring.
更多
查看译文
关键词
Electrocardiography,Field Programmable Gate Array (FPGA),Ensemble Empirical Mode Decomposition (EEMD)
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要