A head-to-head comparison of personal and professional continuous glucose monitoring systems in people with type 1 diabetes: hypoglycaemia remains the weak spot.

DIABETES OBESITY & METABOLISM(2019)

引用 19|浏览23
暂无评分
摘要
To compare the performance of a professional continuous glucose monitoring (proCGM) and a personal continuous glucose monitoring (persCGM) system worn in parallel under standardized conditions in individuals with type 1 diabetes (T1D), two CGM systems (iPro2 - proCGM; Minimed 640G - persCGM) worn in parallel using the same sensor (Enlite 2) were compared. Ten people with T1D were included in this single-centre, open-label study in which CGM performance was evaluated. The study consisted of a 24-hours inpatient phase (meals, exercise, glycaemic challenges) and a 4-day home phase. Analyses included fulfilment of ISO 15197:2013 criteria, mean absolute relative difference (MARD), Parkes Error Grid and Bland-Altman plots. During the inpatient stay, ISO 15197:2013 criteria fulfilment was 58.4% (proCGM) and 57.8% (persCGM). At home, the systems met ISO 15197:2013 criteria by 66.5% (proCGM) and 65.3% (persCGM). No difference of MARD in inpatient phase (19.1 +/- 16.7% vs. 19.0 +/- 19.6; P = 0.83) and home phase (18.6 +/- 26.8% vs. 17.4 +/- 21.3%, P = 0.87) was observed. All sensors performed less accurately during hypoglycaemia. ProCGM and persCGM showed similar performance during daytime and night-time for the inpatient and the home phase. However, sensor performance was reduced during hypoglycaemia for both systems.
更多
查看译文
关键词
clinical trial,continuous glucose monitoring (CGM),hypoglycaemia,type 1 diabetes