An automated intravaginal dynamometer: Reliability metrics and the impact of testing protocol on active and passive forces measured from the pelvic floor muscles.

NEUROUROLOGY AND URODYNAMICS(2018)

引用 13|浏览1
暂无评分
摘要
Aims(1) To determine the reliability of an automated dynamometer designed to assess pelvic floor muscle (PFM) strength and resistance to passive elongation. (2) To evaluate the impact of PFM length and rate of tissue elongation on dynamometric outcomes. MethodsAt each of two sessions, twenty nulliparous women performed three maximum voluntary contractions (MVC) of their PFMs with the dynamometer set to two different anteroposterior (AP) diameters (25mm, 35mm). Next, with PFMs relaxed, the arms of the dynamometer opened three times to 40mm at two speeds (25mm/s, 50mm/s). Outcomes included baseline force, peak force, relative peak force, rate of force development (RFD), stiffness and stress relaxation. Repeated-measures ANOVAs were used to test trial, day, and task effects (=0.05), and intra-class correlation coefficients (ICCs) were computed. ResultsForces measured on MVC were higher with the larger AP diameter, and passive resistance was higher for the faster rate of tissue elongation. The between-trial reliability of all outcomes was excellent (0.82更多
查看译文
关键词
biomechanics,dynamometer,muscle strength (MESH unique ID=D053581),maximal voluntary contraction,pelvic floor (MESH unique ID=D017773)