Comparison of Central Corneal Thickness Measured by Standard Ultrasound Pachymetry, Corneal Topography, Tono-Pachymetry and Anterior Segment Optical Coherence Tomography.

CURRENT EYE RESEARCH(2018)

引用 18|浏览4
暂无评分
摘要
Purpose: To compare central corneal thickness (CCT) measured by standard ultrasound pachymetry (USP), and three non-contact devices in healthy eyes. Methods: A cross-sectional study of CCT measurement in 52 eyes of 52 healthy volunteers was done by a single examiner at Ocular Surface and Contact Lens Laboratory. Three consecutive measurements were done by standard USP, non-contact tono-pachymeter, Pentacam corneal topographer, and Anterior Segment Optical Coherence Tomography (AS-OCT). The mean values were used for assessment. The results were compared using multivariate ANOVA, linear regression, and Pearson correlation. Agreement among the devices was analyzed using mean differences and Bland-Altman analysis with 95% limits of agreement (LoA). Finally, reliability was analyzed using intraclass correlation coefficient (ICC). Results: Mean CCT by ultrasound pachymeter, tono-pachymeter, corneal topographer and AS-OCT were 558.9 +/- 31.2 mu m, 525.8 +/- 43.1 mu m, 550.4 +/- 30.5 mu m, and 545.9 +/- 30.5 mu m respectively. There was a significant positive correlation between AS-OCT and USP (Pearson correlation=0.957, p<0.001), corneal topography and USP (Pearson correlation=0.965, p<0.001), and corneal topography and AS-OCT (Pearson correlation=0.965, p<0.001). There was a lower correlation between CT-1P tono-pachymeter and the other three modalities. Intraclass correlation coefficients show an excellent reliability between pairs except for CT-1P against the other three instruments that were found moderate. Conclusions: CT-1P tono-pachymeter underestimates CCT measurements compared to Scheimpflug system, AS-OCT device, and USP. Mean CCT among USP, Pentacam and AS-OCT were comparable and had significant linear correlations. In clinical practice, these three modalities could be interchangeable in healthy patients.
更多
查看译文
关键词
Central corneal thickness,ultrasound pachymetry,non-contact tono-pachymetry,Scheimpflug camera,anterior segment optical coherence tomography,interchangeability
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要