Using Optical Tracking For Kinematic Testing Of Medical Robots

INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY(2018)

引用 3|浏览4
暂无评分
摘要
BackgroundIn image-guided robotic interventions, an error component is related to the positioning error of the manipulator. Therefore, measuring the kinematic error is required during robot development. However, no specialized measurement device exists for this task. This study explores the possibility of using optical tracking for robot measurement.MethodsA CNC machine is used to position an optical marker, generating a gold standard reference. Repeated position measurements are acquired with an NDI Polaris Hybrid (R) optical tracker at each static location, and averaged. These measurements are compared to the reference set.ResultsAveraging repeated static position measurements improves precision (200-500 samples). Measurement accuracy ranges between 44m and 137m in close proximity of the tracker.ConclusionsRepeated static position measurements in the near field of view enable the optical tracker to outperform its general-purpose accuracy specification. Optical tracking may be used to test robot kinematics with a high degree of accuracy.
更多
查看译文
关键词
NDI Polaris, optical tracker, robot kinematic tests