Quantification of dehydroepiandrosterone in human serum on a routine basis: development and validation of a tandem mass spectrometry method based on a surrogate analyte

Analytical and bioanalytical chemistry(2017)

引用 9|浏览52
暂无评分
摘要
In the clinical laboratories, dehydroepiandrostenedione (DHEA) is usually quantified by immunoassay-based methods, which are often affected by cross-reactivity with endogenous interferences, such as 4-androsten-3β-ol-17-one. The interfering compounds lead to a poor accuracy of the measurements, mainly at a low concentration level. The present paper describes a validated method based on tandem mass spectrometry coupled to liquid chromatography, for the accurate quantification of DHEA in serum. The peculiarity of this method is the use of calibrators and quality controls prepared by adding measured amounts of DHEA-D 5 , a stable isotope-labeled analogue of DHEA, to real serum from healthy subjects. DHEA-D 5 is used in place of DHEA, which is usually present in unstripped serum at physiological levels, as it has the same basic structure, provides an equivalent instrumental response, and can be easily distinguish by DHEA by mass spectrometry due to its different m / z value. The method proved to be sensitive, with a LLOD of 0.09 ng/mL and a LLOQ of 0.23 ng/mL, and selective, with overall performances that allow its use on a routine basis.
更多
查看译文
关键词
DHEA,DHEA-D5,Laboratory medicine,Liquid chromatography,Tandem mass spectrometry