Proof of concept testing of a positive reference material for in vivo and in vitro skin irritation testing.

JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS(2018)

引用 3|浏览15
暂无评分
摘要
In vivo and in vitro irritation testing is important for evaluating the biological safety of medical devices. Here, the performance of positive reference materials for skin irritation testing was evaluated. Four reference standards, referred to as Y-series materials, were analyzed: a polyvinyl chloride (PVC) sheet spiked with 0 (Y-1), 1.0 (Y-2), 1.5 (Y-3), or 10 (Y-4) parts of Genapol X-080 per 100 parts of PVC by weight. Y-1, Y-2, and Y-3 did not induce skin irritation responses in an in vitro reconstructed human epidermis (RhE) tissue model, as measured by tissue viability or interleukin-1 release, or in an in vivo intracutaneous response test using rabbits. In contrast, Y-4 extracts prepared with saline or sesame oil at 37 degrees C and 50 degrees C clearly elicited positive irritation responses, including reduced viability (< 50%) and significantly higher interleukin-1 release compared with the solvent alone group, in the RhE tissue model and an intracutaneous response test, where substantial necrosis was observed by histopathology. The positive skin irritation responses induced in vitro under various extraction conditions, as well as those elicited in vivo, indicate that Y-4 is an effective extractable positive control material for in vivo and in vitro skin irritation tests of medical devices. (c) 2017 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 106B: 2807-2814, 2018.
更多
查看译文
关键词
irritation test,positive control,biological safety evaluation,guidance,non-ionic surfactant
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要