Establishing an Accuracy Basis for the Vitamin D External Quality Assessment Scheme (DEQAS).

JOURNAL OF AOAC INTERNATIONAL(2017)

引用 41|浏览25
暂无评分
摘要
Until recently, the Vitamin D External Quality Assessment Scheme (DEQAS) assessed the performance of various assays for the determination of serum total 25-hydroxyvitamin D [25(OH)D] by using a consensus mean based on the all-laboratory trimmed mean (ALTM) of the approximately 1000 participants' results. Since October 2012, the National Institute of Standards and Technology (NIST), as part of the Vitamin D Standardization Program, has participated in DEQAS by analyzing the quarterly serum sample sets using an isotope dilution LC-tandem MS (ID LC-MS/MS) reference measurement procedure to assign an accuracy-based target value for serum total 25(OH)D. NIST has analyzed 90 DEQAS samples (18 exercises x 5 samples/exercise) to assign target values. The NIST-assigned values are compared with the ALTM and the biases assessed for various assays used by the participants, e.g., LC-MS/MS, HPLC, and several ligand-binding assays. The NIST-value assignment process and the results of the analyses of the 90 DEQAS samples are summarized. The absolute mean bias between the NIST-assigned values and the ALTM was 5.6%, with 10% of the samples having biases >10%. Benefits of the accuracy-based target values are presented, including for sample sets with high concentrations of 25(OH)D-2 and 3-epi-25(OH)D-3.