Validation of temporal gait metrics from three IMU locations to the gold standard force plate.

2016 38th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC)(2016)

引用 12|浏览26
暂无评分
摘要
The purpose of this work is to compare temporal gait parameters from three different IMU locations to the gold standard force platform. 33 subjects (12 F, 21 M) performed twenty gait trials each while wearing inertial measurement units (IMUs) on the trunk, both shanks and both feet. Data was simultaneously collected from a laboratory embedded force plate. Step times were derived from the raw IMU data at the three IMU locations using methods that have been shown to be accurate. Step times from all locations were valid compared to the force plate. Foot IMU step time was the most accurate (Pearson = .991, CI width = 3.00e2), the trunk IMU was the next most accurate (Pearson = .974, CI width = 4.85e2) and shank step time was the least accurate (Pearson = .958, CI width = 6.80e2). All three sensing locations result in valid estimations of step time compared to the gold standard force plate. These results suggest that the foot location would be most appropriate for clinical applications where very precise temporal parameter detection is required.
更多
查看译文
关键词
Adult,Algorithms,Female,Gait,Humans,Lumbar Vertebrae,Male,Walking,Young Adult