Surface characterisation and electrochemical stability of anodised new alloy in simulated physiological electrolytes

MATERIALS AND CORROSION-WERKSTOFFE UND KORROSION(2014)

引用 1|浏览2
暂无评分
摘要
In this paper, we applied the potentiostatic anodisation method on the new Ti-20Nb-10Zr-5Ta alloy surface with the aim to increase its stability and bioactivity. The morphology (SEM) and composition (Raman, FT-IR) of the obtained anodisation nanolayer are studied. Also, electrochemical stability (by cyclic potentiodynamic polarisation and EIS), corrosion resistance (by linear polarisation), long-term behaviour (by monitoring of the open circuit potentials and corresponding open circuit potential gradients) and the characterisation of the layers deposited on the alloy surface after 1000 h in acid, neutral and alkaline Ringer and neutral Ringer-Brown solutions are presented. By potentiostatic anodisation, on the alloy surface was electrodeposited a continuous nanolayer containing TiO2 rutile type oxide. The anodised layer possesses better protective properties and at the same time better porosity, therefore a higher stability and bioactivity in comparison with the native passive film. After 1000 h in the testing solutions promoted the deposition of the brushite, phosphatotitanate and hydroxyapatite, precursors, respectively component of the human bone, namely this layer is bioactive.
更多
查看译文
关键词
sem,raman,xrd,eis
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要