Optimization of acquisition parameters and accuracy of target motion trajectory for four-dimensional cone-beam computed tomography with a dynamic thorax phantom

Radiological physics and technology(2014)

引用 8|浏览1
暂无评分
摘要
Our purpose in this study was to evaluate the performance of four-dimensional computed tomography (4D-CBCT) and to optimize the acquisition parameters. We evaluated the relationship between the acquisition parameters of 4D-CBCT and the accuracy of the target motion trajectory using a dynamic thorax phantom. The target motion was created three dimensionally using target sizes of 2 and 3 cm, respiratory cycles of 4 and 8 s, and amplitudes of 1 and 2 cm. The 4D-CBCT data were acquired under two detector configurations: “small mode” and “medium mode”. The projection data acquired with scan times ranging from 1 to 4 min were sorted into 2, 5, 10, and 15 phase bins. The accuracy of the measured target motion trajectories was evaluated by means of the root mean square error (RMSE) from the setup values. For the respiratory cycle of 4 s, the measured trajectories were within 2 mm of the setup values for all acquisition times and target sizes. Similarly, the errors for the respiratory cycle of 8 s were <4 mm. When we used 10 or more phase bins, the measured trajectory errors were within 2 mm of the setup values. The trajectory errors for the two detector configurations showed similar trends. The acquisition times for achieving an RMSE of 1 mm for target sizes of 2 and 3 cm were 2 and 1 min, respectively, for respiratory cycles of 4 s. The results obtained in this study enable optimization of the acquisition parameters for target size, respiratory cycle, and desired measurement accuracy.
更多
查看译文
关键词
Optimization, Four-dimensional computed tomography, Respiratory motion, Image-guided radiation therapy
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要