Three dimensional drop tracking flow chamber for coalescence studies.

Anne M Grillet,Carlton F Brooks, Chris J Bourdon, Allen D Gorby

REVIEW OF SCIENTIFIC INSTRUMENTS(2007)

引用 1|浏览13
暂无评分
摘要
We have developed a novel flow chamber which imposes a controlled axisymmetric stagnation flow to enable the study of external flow effects on coalescence dynamics. This system allows for the first time the precise positioning of a drop in a three dimensional flow and additionally enforces a highly symmetric flow around the drop. We focus on the study of a single drop approaching a stationary flat plane as this is analogous to two drops approaching each other. A single drop is created and then guided along the unsteady center line of a stagnation flow. The real time computer control algorithm analyzes video images of the drop in two orthogonal planes and manipulates flow restricting valves along the four outlets of the flow. We demonstrate using particle image velocimetry that the computer control not only controls the drop position but also ensures a symmetric flow inside the flow chamber. This chamber will enable a detailed investigation of the drainage of the thin film between the drop and the lower surface in order to probe the effect of external flow on coalescence. (c) 2007 American Institute of Physics.
更多
查看译文
关键词
thin film,real time systems,flow control,particle image velocimetry,three dimensional,real time computing