Real Time Breathing Rate Estimation From A Non Contact Biosensor

2010 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC)(2010)

引用 14|浏览1
暂无评分
摘要
An automated real time method for detecting human breathing rate from a non contact biosensor is considered in this paper. The method has low computational and RAM requirements making it well-suited to real-time, low power implementation on a microcontroller. Time and frequency domain methods are used to separate a 15s block of data into movement, breathing or absent states; a breathing rate estimate is then calculated. On a 1s basis, 96% of breaths were scored within 1 breath per minute of expert scored respiratory inductance plethysmography, while 99% of breaths were scored within 2 breaths per minute. When averaged over 30s, as is used in this respiration monitoring system, over 99% of breaths are within 1 breath per minute of the expert score.
更多
查看译文
关键词
microcontroller,impedance measurement,biosensors,real time,frequency domain analysis,real time systems,frequency domain,microcontrollers,estimation,patient monitoring,plethysmography,bluetooth,sleep