Assessment of intra- and inter-laboratory variation in multi-centre real-time reverse-transcribed PCR (qRT-PCR)-based mRNA quantification of a prognostic marker in breast cancer using external quality assurance (EQA).A questionnaire on the methodologies used and EQA calibrators were sent to 5 participating laboratories from 4 European countries, which measured mRNA levels of PITX2 splice variants and reference genes by qRT-PCR.Differences in the methodology included PCR quantification methodology and equipment, RNA extraction and cDNA synthesis procedures. The intra-laboratory coefficient of variation (CV) ranged from 5 to 23%, and the inter-laboratory CV ranged from 17 to 30%. The inter-laboratory CV was reduced to 13% by using prediluted calibrators and by harmonising the data in the central QA laboratory. Additional normalisation using reference genes did not decrease the variation further.Both externally provided calibrators and centralised harmonisation are required to reduce the intra-laboratory variation in multi-centre qRT-PCR results to an acceptable level.