Noninvasive Ultrasound Imaging for Bone Quality Assessment Using Scanning Confocal Acoustic Diagnosis, muCT, DXA Measurements, and Mechanical Testing.

ICMB'08: Proceedings of the 1st international conference on Medical biometrics(2008)

引用 1|浏览12
暂无评分
摘要
Osteoporosis is a disease characterized by decreased bone mass and progressive deterioration of the microstructure, affecting both mineral density and bone's fragility. Current diagnoses are only measuring apparent bone mineral density (AppBMD). Using our newly developed scanning confocal acoustic diagnostic (SCAD) system, we evaluated the ability of quantitative ultrasound in noninvasively predicting bone's quantity and quality on 19 human cadaver calcanei. Results show that ultrasound attenuation image on intact calcaneus represents bone mass distribution. High correlation (R=0.82) exists between SCAD determined broadband ultrasound attenuation (BUA) and DXA determined AppBMD at the calcaneus, as well as in the AppBMD result at femoral neck (R=0.81). SCAD determined BUA and ultrasound velocity (UV) are highly correlated with the micro-CT and mechanical testing determined bone quantity and quality parameters. These results suggest that image-based quantitative ultrasound is able to identify ROI and predict both bone mass and strength.
更多
查看译文
关键词
bone mass,apparent bone mineral density,bone mass distribution,bone quantity,broadband ultrasound attenuation,image-based quantitative ultrasound,quantitative ultrasound,ultrasound attenuation image,ultrasound velocity,AppBMD result,DXA measurement,Noninvasive ultrasound imaging,bone quality assessment,confocal acoustic diagnosis,mechanical testing