Interaction of force-fitting and surface roughness of implants.

R Skalak,Y Zhao

Clinical implant dentistry and related research(2000)

引用 37|浏览6
暂无评分
摘要
Increased surface roughness may increase installation torque and thus appear to increase the initial stability of an implant. However, it is not immediately clear if the increased torque is attributable to an increase in the effective diameter of the implant or to increased resistance of the bone because of the greater roughness.Force-fitting stresses arise when an implant is placed into a predrilled hole of smaller-diameter in bone. The purpose of this report is to discuss the interaction of force-fitting stresses and surface roughness effects and to develop some general guidelines as to clinical procedures based on this theory.Solutions for the force-fitting stresses are derived from well-known equations of elasticity.Substantial force-fitting stresses on the order of several tens of MPa can be generated when a titanium cylinder is placed into a hole in bone, the diameter of which is only 100 microns smaller.When a hole slightly smaller than the implant diameter is prepared for implant placement, force-fitting stress increases installation torque and stability can be induced. Thus, large surface roughness of implants should not be viewed as an exclusive mechanism for providing a desirable level of initial fixity. Smaller roughness with the same mean diameter is equally effective.
更多
查看译文
关键词
surface roughness,osseointegration
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要