SU-FF-T-20: Near Source Dosimetry of HDR Base of Tongue Treatments Using BrachyVision TPS & GrafChromic HD-810 Film

MEDICAL PHYSICS(2009)

引用 0|浏览4
暂无评分
摘要
Purpose: To determine the accuracy of the BrachyVision HDR treatment planning computer at predicting near source doses in the presence and absence of bone scatter in base of tongue treatments. Materials & Methods: A HDR treatment plan was developed to deliver a uniform planer dose at a depth of 1cm from a custom made planer geometry applicator with 5 catheters separated by 1cm. Plan dwell times were scaled to deliver doses of 10, 20, 50, 75, and 150Gy for a dose calibration curve for GrafChromic HD‐810 film. Dose measurements were performed at distances of 0mm, 1mm, 2mm, 3mm, 5mm, 7mm, & 10mm from the surface of the catheters. Three prescription doses were used for these measurements depending on distance from the catheters; 10Gy (0mm), 15Gy (1mm), & 25Gy (2, 5, 7, 10mm). Measurements were performed in solid water and in solid water directly against and 1mm away from a bone equivalent backscatter (Schedule 80 PVC). Doses were compared using 35 point doses (7 along each of 5 catheters) at 1cm increments. Results: Analysis of the films showed average percent differences of; 24±13%, ∼0±6%, 4±4%, 2±6%, 2±6%, & 4±3% (0, 1, 2, 5, 7, 10mm respectively) for solid water, 31%, 5%, 9%, 11%, 16%, & 14% (0, 1, 2, 5, 7, 10mm respectively) with similar standard deviations to solid water, for directly against the PVC. Conclusions: For this planar geometry are within a standard deviation of predicted doses for a homogeneous media except directly against the catheter. At a plane 5mm from the catheter the presence of high‐z back scatter produced an average increase in dose of 5%.
更多
查看译文
关键词
tongue treatments,brachyvision tps,dosimetry,hdr base,su-ff-t
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要